Techne Precision Fluidized baths 
Request literature on this product via mail
Check out the YouTube video below on this bath

The FB08 Fluidized bath offers excellent temperature stability and uniformity making it a good choice for temperature sensor comparison calibration work and shape setting of metals including Nitinol and platinum for Medical device manufacturing. With the optional probe holder which acts as an isothermal block, stabilities of better than +/-0.008ºC in dead-bed mode can be realized. Combine the FB08 with one of our Platinum resistance thermometer probes and indicator for a high performance temperature system.  Here are some of the benefits of the FB08 Fluidized bath:

bullet

The Fluidized bath of choice for critical processes requiring the highest stability and uniformity at temperatures up to 700ºC/1292ºF.

bullet

Fluidized bath action is non-abrasive to any devices or parts put into the bath including those highly polished.

bullet

Much more efficient than ovens (all heat is lost when opening the door) and salt baths. Fast recovery time after the bath is quenched.

bullet

Much Safer than volatile salt baths

bullet

Short term (15 minutes) stability and uniformity typically better than +/-0.2ºC

bullet

Typical thermal profile of the FB08 (164kB PDF)

bullet

FB08 FAQ's (29kB PDF)

bullet

Download the FB08 instruction manual (259kB PDF)

bullet

Download the Brochure (160kb) 

Techne FB08 Precision Fluidized bath



Shape setting of Nitinol Medical Stents


Test and calibration of temperature
 sensors and systems


Description
:
A rectangular bench standing unit, the controls for temperature and fluidizing air flow are mounted on a recessed panel on the front, together with on-off switches for the power supply and the dust extractor and indicators for various functions. Insulated and vented the outer case is safe to touch even at the maximum operating temperature. When Fluidized, the media is heated by four immersion heaters close to the container wall with the control thermocouple between the heaters. The internal components are protected by an air cutout switch that disables heat in the event the fluidizing air is lost.
 
A clean/dry inert gas from an external source passes through two filters and regulators, to reduce the pressure for the cyclone extraction system and fluidization. Fluidizing air passes through a flow meter and then to the plenum chamber from which it is distributed evenly around the inner container. Dust extraction is by means of ambient air drawn from around the probe holder/plate and then through a cyclone to the exhaust filter. Alumina dust removed from the opening of the bath is deposited into the front mounted collection jar where it can be removed and dumped back into the bath.


Operation: Setting of the fluidizing air is by means of a combined control valve and flow meter. (Note: air adjustment is automatic with the FB08C) Fluidizing air requirements are inverse of the bath temperature. i.e- lower bath temperatures require more air for proper fluidization. The bath temperature is set by means of a front panel mounted PID controller. Sensors for calibration should be suspended as close as possible to the center of the bath. When calibrating sensors the probe holder is recommended as it is custom built with the dimensions of the sensors you will be calibrating.


Dead-bed method

A major feature of the FB08 series is calibration using the Dead-bed method. During Dead-bed the power is shut off and 30 seconds later the fluidizing and extraction air are turned off. This causes the alumina to collapse around your work and provides a dense, high thermal insulation environment. For a period of three to ten minutes thermal equilibrium occurs and stability of better than ±0.008°C is observed.

Dead bed stability

Models to choose from:
The FB08 is offered in three different models. The standard FB08 operates from +50 to 700°C, in addition to the standard unit the FB08C offers automatic fluidizing air control and is capable of remote PC operation. When connected to your liquid nitrogen supply the FB08LT operates between -100 and +200°C. The single probe plate and ring are included with the baths. You will also receive one 35lb. charge of alumina to get started.

Specifications

 

Specification

FB08 FB08C FB08LT
Temperature range °C 50 to 700 50 to 700 -100 to +200
Temperature Stability      

@500°C over a 5 minute period 

±0.1 ±0.1 ±0.08 at 200°C
Dead-bed stability ±0.008 ±0.008 ±0.01
Display Resolution 1 1 1
Type of Control 3 term PID 3 term PID 3 term PID
Sensor Type K' thermocouple K' thermocouple 100 ohm PRT
Heat up time, minutes      
20 to 700°C 105 105  
20 to 200°C     30
Cooling time, minutes      
700 to 200°C 165 165  
200 to -100°C     90
Air Pressure (PSI) 60 60 60
Maximum flow liters/minute 127 127 170
Weight of medium, lbs. 35 35 35
LN2 consumption maximum liters/hour n/a n/a 7
Overall size LxWxH, inches 30.3 x 20.3 x 23.6 34.3 x 20.3 x 23.6 30.3 x 24 x 23.6
Working diameter x depth, inches 6.5 x 15 6.5 x 15 6.5 x 15
Warranty 1 year parts and labor
Catalog number 3021400 3022100 3021600

Accessories
Click on an picture to see it larger

Image

Description

Catalog Number

Basket probe/plate 6037759
Probe carrier/holder (specify probe diameters)
Use for comparison calibrations of temperature sensors and systems. Stability and uniformity better than 0.010°C in dead bed mode.
7031993
Probe plate ring
(included with the bath)
6007763
Probe plate
(included with the bath)
6007764
Air pressure regulator/filter 6035915

Replacement Spare parts

Description

Catalog Number

Jar 6007557
8 micron filter 6007558
Jar Seal 6007559
1 (35lb.) charge of alumina 7030794
15 amp fuse 6008181
Porous plate 6001557
Heater 6007477
Air mover sleeve 6007359
Air mover Jetflow 6006985

 

up.gif (1352 bytes)

Home | Temperature Calibration | Online Store | News
Contact Us | About Us
Copyright © 2008 Techne® Incorporated