|
Thermal
Cyclers
I bought a new Techne Thermal
Cycler and need help in changing the fuses.
For directions
click here
How long has Techne been making
Thermal Cyclers?
Experience in thermal cycler manufacturing since 1987, ensures that
Techne Thermal Cyclers provide guaranteed performance and quality results,
every time.
Can I transfer protocols from other
brands to Techne thermal cyclers?
Yes, all Techne cyclers have adjustable ramp speed.
I
have transferred a protocol from another brand of Thermal Cycler to a
Techne Thermal Cycler. Why am I not getting a good yield from my Techne
unit?
There could be
several possibilities. One important thing to remember is that
there could be a slight temperature difference between any two Thermal
Cyclers. Optimizing temperature and hold time may be necessary. Most of
the time, adjusting the ramp speed (temperature characteristics: heating
and cooling rates) on the Techne Thermal Cycler to match the ramp speed of
your older unit would resolve the problem.
Do I need to have a final 4°C
hold temperature?
Most Thermal Cyclers use
peltier elements to heat and cool the block. To extend the life of
your block we recommend a final hold temperature above 10°C
or disabling the final hold segment especially when holding over night or
weekends. As peltiers are solid state devices, they are prone to
drift with age which results in block non-uniformity.
Techne can perform a Thermal cycler
thermal validation on any make or model to gain some early insight into
possible problems. Visit our calibration page at http://www.techneusa.com/support/thermalcycler.htm
or for more details on this service!
Can TC-512 be used to do
melting point analysis?
To generate melting point curves, you need real time PCR
machine. TC-512 is a gradient cycler. It doesn’t have the ability to do real
time PCR
For In-situ
PCR should the heated lid be disabled?
In-situ PCR is performed on a special heating block that is
capable of holding glass slides. It depends on your protocol and it’s up to do
if you would like to enable or disable the heated lid. For most in-situ PCR,
heated lid is turned off.
TC512 was used to run a 12 degree gradient to find the
annealing temperature of a tricky primer. The
annealing temperature from running a gradient was found to be 59.7 C; the PCR
was run by setting the annealing temperature to 60C. The result didn’t show any
bands, why? What should I do?
There are several possibilities. One thing that might have
caused this type of result could be that you forgot to add the primer or an
important enzyme. Another thing could be that the concentration is not the
same.
0.3C degrees could make a lot of difference if you are doing a long PCR.
What you should do now is run the PCR again carefully adding all the enzymes.
If you see the same results, then run another gradient (tighter 4C gradient) and
use that annealing temperature to run your PCR.
I am using a non skirted plate (96x.2ml) to run my PCR reaction in
TC412. The volume in certain places is 20 ul and in others it is 25 ul. Would
this difference in volume make a difference in the PCR reaction? Also, do I
need to enter the volume somewhere in the program?
No, the different volume in the plate would not make a
difference in your PCR reaction. You should always use half skirted or non
skirted plates with Techne’s Thermal Cyclers to get a good PCR result. Full
skirted plate would not give you a good PCR result. Any PCR plate would work
with Techne’s cyclers. We recommend you use Axygen’s PCR tubes and
plates.
I am doing a PCR reaction and noticed sample
condensation at the top of the tube. Please help
There are several possibilities. First, check that the lid
is adjusted down on top of the tubes. If not then turn the lid clockwise to
adjust it down on top of the tubes. Second, make sure you have “preheat lid”
option enabled. You can check this by viewing your program. If you are having
trouble, please contact us at 1-800-225-9243 (313)
I am trying to
decrement the temperature 0.5°C
for 20 cycles. How should I do it?
The procedure is very simple.
Step 1: Highlight the segment and press enter
Step 2: Press pause; prompt line will change between change seg?
Yes/No
Step 3: Press enter to choose yes
Step 4: Press up arrow; (the seg will change to fst) edit your temperature,
press enter and then come down by pressing the down arrow
Step 5: The second line would read lst; edit the temperature and time and press
enter; fst meaning first and lst meaning last.
The final result would like this
Seg Temperature Time Ramp
Speed
fst 68°C
0m30 max
lst 58°C 0m30
I am getting an error message on my
Thermal Cycler screen. What should I do?
Please note down the error message and call our service department
(1-800-225-9243 ext. 373) to resolve the problem.
Spectrophotometers
My Genova Life Science Analyzer is
experiencing an error message during start-up; what do I do?
Error message is "operating parameter failure" - Simple press "enter" and
the instrument will continue test
Error message is "dark level test" - if lid is open during start up test, this
message will appear. Simply close the lid and re-start.
Error message is "wavelength calibration failure" - there are actually a few
simple reasons why this will appear:
a: Lid is opened during test - close lid and restart
b: Cuvette is in cuvette holder- take out cuvette, close lid and restart
c: Platform holding cuvette holder is loose or not properly aligned - this could
have occurred in shipping. The platform is held in place by a large
thumbscrew on the underside of the cuvette chamber. Loosen screw, ensure
that the platform is sitting tightly and evenly and tighten screw. Close
lid and restart.
IMPORTANT - Ensure that nothing is in sample chamber and lid is closed during
start- up
If this does not help, then it needs to be serviced. Please call
800-225-9243 (374)
Why am I getting an error
message (error 2) / low light level in the UV mode when measuring OD.
The 100 cuvettes that come with the spectrophotometers are visible only. In
other words those cuvettes can measure only at the visible range, if you try to
measure @ the UV range you will get error 2 message. Please use the
appropriate cuvette, for cuvette information please call 800-225-9243 (313)
My Genova display screen is hard to read; how can I adjust the contrast?
Go to the main Genova menu and press the left or right direction arrows for
contrast adjustment.
How do I measure 260/280 ratio on 6300 and get the
answer in concentration.
First calibrate the unit at 260 and measure the absorbance. Then, calibrate the
unit at 280 and measure the absorbance. Use these absorbance values and plug
them into the equation to calculate the concentration.
Without a reference
wavelength: ((Abs1 x factor 1) - (Abs2 x factor 2)) x dilution
Using a reference wavelength (320nm): (((Abs1 - Abs ref) x factor 1) - ((Abs2 -
Abs ref) x factor 2)) x dilution
Hybridization Ovens
The mini tubes used for blotting techniques needs to be
autoclaved at 300 C. Can I autoclave them?
Yes, the tubes can be autoclaved (please take the plastic caps off before you
autoclave the tubes; plastic caps can not be autoclaved)
I want to know if the HB-1D is capable of both northern blotting and in situ hybridization.
Yes, HB-1D can be used for in situ hybridization. In addition to in situ
hybridization, it is used for Southern blots, Northern blots, Western blots, Dot
blots, and Slot blots. The temperature range for HB-1D is 10C above ambient to
100C.
What is the
temperature stability around 65C on HB-1D.
The stability (in tubes) across the entire range (10C above ambient to 100C) is
<± 0.1°C. HB-1D features an extremely uniform and stable temperature
environment. The insulated door prevents heat loss and features a large double
glass observation panel for viewing the chamber contents.
How do you control the speed?
There is a rotation control speed knob for controlling the
speed. 0, 5 -20 oscillation per minute. On older models HB-1D, the
rotation speed is fixed.
Why is the minimum temperature 10°C above
ambient?
It is for the internal heat of the unit. When the unit is turned on it
generates heat.
Why should I use a separation membrane?
It is recommended that you place a separation membrane if placing more than
one blot in a bottle. However, if you are placing one blot per bottle, then you
don’t need a separation membrane.
How effective are the separation membranes and is it
necessary to use them?
The separation membranes will eliminate the blots from coming in contact with
each other which causes background black spots that are not removable.
Cell Culture
Equipments
Can I autoclave the culture vessels?
Yes, the vessels can be autoclaved.
Miscellaneous
Why
should we choose
Techne?
Techne offers you the flexibility you need with the quality you
expect from an established world leader in temperature control instrumentation.
Our company goal is to provide maximum benefit to
our customers; our more than 57 years of successful history proves it. Our
strongly motivated customer support team offers outstanding service and best
available products to the customers.
I would like to become an OEM, original equipment
manufacturer, customer; what should I do?
OEMs are not the original manufacturers -- they are the customizers of a
particular product. You can either get the product with Techne’s label
(customize it on your own) or with a customized label (done by Techne) with your
company’s name/logo on it. There is a minimum order of 10 units. You can do a
blanket order of 10 and the units will be shipped according to the need.
Please call us for more information or assistance.
|