Model 6310 Scanning Spectrophotometer
Application Notes
Request literature on this product via mail

  • Photometric measurements in Absorbance and Transmittance

  • Concentration measurement by direct factor entry or against blank and standard

  • Quantitation using up to 6 standards - with curve correction and correlation coefficient

  • Kinetics with real time graphical display of sample and standard runs

  • Spectrum - scanning any range from 320-1000nm

  • Store up to 50 methods in each mode

  • Includes free 63Ten  PC software

    An advanced measuring system with modes for Photometrics, Concentration, Quantitation, Kinetics and Scanning. The 6310 will meet the demands of a wide range of applications, especially those in the clinical, veterinary, environmental, general QC and performance testing areas.

Features include:

With a memory capacity of 50 methods for each mode, the fully open operating system ensures compatibility
 with reagent kits from most manufacturers, as well as enabling full customization to the operator's choice of
chemistries and operating procedures.

Simple keypad layout and novel customizable menus ensure reliable operation by unfamiliar users.

Supervisor Security functions, when activated, enable all the set up options to be locked. These cannot be
changed without the entry of a security code number.

Simultaneous display of Absorbance, Transmission and Wavelength in the Photometrics mode means that
manual range changing is not necessary.

Simple concentration measurements using a single standard or factor can be made in the concentration
mode. The Quantitation mode allows more complex concentration methods to be programmed, using up
 to 6 standards to create a calibration curve. For each curve fit option a linearity correlation coefficient
is calculated.

A real-time graphical display of Kinetic measurement is shown, along with the calculated concentration
value and other relevant data.

The Spectrum Scanning mode allows scans to be made across and selected portion of the 320-1000nm
range. The auto-ranging Absorbance axis allows scans to be commenced with no prior knowledge of
the peak Absorbance level expected.

The 6310 has an analogue output that can be used to drive chart recorders and other analogue recording
 or conversion devices.

A bi-directional RS232 serial port allows connection to serial printers and computers.

The  63Ten software allows full emulation of the instrument functions on a PC, including the
importing of results and download of methods. The results can be printed out to any attached windows printer.

An electrically heated cuvette holder and control unit is available, as well as a water heated cuvette
holder and a thermostatically controlled water circulator. The external,programmable sipper pump
ensures faster and safer sample handling.

The large sample compartment enables a wide range of sampling devices to be fitted. Cuvette holders
from 10 to 100mm in path length, a 4 position rotary cuvette holder and test tubes from 12.5 to 25mm
can all be accommodated.

Model 6310 - Specifications

Optics

Single beam

Wavelength range

320-1000nm

Resolution

1nm

Accuracy

±2nm

Bandwidth

8nm

Transmittance range

0 to 199.9%

Resolution

0.1%

Accuracy

±1%T

Absorbance range

-0.300 to 1.999A

Resolution

0.001A

Concentration range

-300 to 1999

Selectable resolution

1/0.1/0.01/0.001

Calibration

Blank with a single standard or factor

Memory

50 methods

Quantitation

-300 to 1999Conc

Selectable resolution

1/0.1/0.01/0.001

Calibration

Blank with up to 6 standards

Curve Correction

Linear regression, interpolation or linear regression through zero

Memory

50 methods

Kinetics

Graphical and calculated concentration value

Calibration

Against standard and factor

Correlation factor

r squared value displayed with result

Memory

50 methods and 5 graphed results

Spectrum

Any range between 320 and 1000nm

Scan data interval

1, 2 or 5nm selectable

Analysis

Absorbance and wavelength of peaks and valleys

Memory

50 methods and 5 scans

GLP

Real time clock and calendar
Operator ID
Supervisor security (locks all set up parameters)
Method display lock

Photometric noise levels

<1%

Photometric stability

Better than 1%/hr

Readout

Custom LCD

Outputs

Analogue / RS232 serial port

Input voltage

115/230V a.c. -20% +10% 50/60Hz

Input power

<50W

Size (inches)

14 x 11 x 6

Weight

6Kg

Model 6310 - Order Codes

631 001

Model 6310 Visible range scanning spectrophotometer (320-1000nm) supplied with mains lead, pack 100 disposable cuvettes, 10x10mm cell holder, 63Ten Software package and instruction manual. For use on 230V/50Hz

631 031

Model 6310 Visible range scanning spectrophotometer (320-1000nm) supplied with US line cord, pack 100 disposable cuvettes, 10x10mm cell holder , 63Ten Software package and instruction manual. For use on 110V/60Hz

Model 6310 - Accessories

633 001 Heated cuvette system (cuvette holder and external control unit) supplied with interconnection cables, mains cables for EU, UK and USA and instruction manual. (Universal voltage and frequency)
632 001 External Sipper Pump supplied with inlet and outlet tubing, mains cable and instruction manual (230V/50Hz)
632 031 External Sipper Pump supplied with inlet and outlet tubing, mains cable and instruction manual (110V/60Hz)
630 005 Adjustable path length cuvette holder (10-100mm)
630 204 10mm cuvette holder (supplied as standard with product)
630 020 Test tube holder (13mm diameter)
630 021 Test tube holder (25mm diameter)
630 022 Test tube block (16mm diameter)
634 001 4 position rotary cuvette holder
648 001 Water heated cuvette holder
037 201 Thermostatted water circulator 230V
037 202 Thermostatted water circulator 110V
7062828    Spectrophotometer Printer
543 501     Portable Thermal Printer
060 287 Paper roll for printer
060 288 Printer ribbon
542 009 Interface cable kit
021 041 DC/AC power converter - enables the product to be powered from a 12v DC supply
630 028 Dust cover
035 088 Calibration set
033 290 Storage case
7062828 Spectrophotometer Printer supplied with cable

Glass Cuvettes             Please click here to view images

Jenway Cat. # Quantity Footprint (mm) UV/visible Material   Volume Masked or
non-masked
Wavelength (nm) Path Length (mm) Path Height (mm)
Glass Cuvettes
035-123 1/pk 10x10 Visible Glass 500ul non-masked 320-2500 10 15
035-126 1/pk 10x10 Visible Glass 1ml non-masked 320-2500 10 15
035-027 1/pk 10x10 Visible Glass 3ml-4.5ml non-masked 320-2500 10 15
035-086 1/pk 20x10 Visible Glass 7ml non-masked 320-2500 20 15
035-029 1/pk 40x10 Visible Glass 12-18ml non-masked 320-2500 40 15
035-087 1/pk 50x10 Visible Glass 17.5ml non-masked 320-2500 50 15
035-079 1/pk 100x10 Visible Glass 35ml non-masked 320-2500 100 15
Glass Flow Cells
035-025 1/pk 10x10 Visible Glass 1.8ml non masked 320-2500 10 15
035-045 1/pk 10x10 Visible Glass 80ul non-masked 320-2500 10 15

Quartz Cuvettes    Please click here to view images

Jenway Cat. # Quantity Footprint (mm) UV/visible Material   Volume Masked or
non-masked
Wavelength (nm) Path Length (mm) Path Height (mm)
Quartz Cuvettes
035-124 1/pk 10x10 UV-Visible Quartz 50ul Masked 200-2500 10 15
035-138 1/pk 10x10 UV-Visible Quartz 100ul Masked 200-2500 10 15
035-139 1/pk 10x10 UV-Visible Quartz 200ul Masked 200-2500 10 15
035-125 1/pk 10x10 UV-Visible Quartz 500ul non-masked 200-2500 10 15
035-127 1/pk 10x10 UV-Visible Quartz 1ml non-masked 200-2500 10 15
035-028 1/pk 10x10 UV-Visible Quartz 3.5ml Masked 170-2700 10 15
Quartz Flow Cells
035-044 1/pk 10x10 UV-Visible Quartz 1.8ml non-masked 190-2700 10 15
035-047 1/pk 10x10 UV-Visible Quartz 80ul masked 200-2500 10 15

Plastic Disposable Cuvettes and glass test tubes

Jenway Cat. # Quantity Footprint (mm) DNase and RNase free UV/visible Material   Volume Masked or
non-masked
Wavelength (nm) Path Length (mm) Path Height (mm)
Plastic Disposable Cuvettes
7062810 100/pk 10x10 - UV-Visible Plastic 70ul non-masked 220-900 10 15
7062812 100/pk 10x10 Yes UV-Visible Plastic 70ul non masked 220-900 10 15
060-087 100/pk 10x10 - Visible Plastic 1.5-3ml non-masked 340-900 10 15
035-132 100/pk 10x10 - UV-Visible Plastic 1.5-3ml non-masked 200-900 10 15
060-231 100/pk 10x10 - UV-Visible Plastic 1.6ml non-masked 200-900 10 15
060-084 100/pk 10x10 - Visible Plastic 2.4-4.5ml non-masked 340-900 10 15
060-229 500/pk 10x10 - Visible Plastic 2.4-4.5ml non-masked 340-900 10 15
060-230 100/pk 10x10 - UV-Visible Plastic 2.4-4.5ml non-masked 280-1100 10 15
060-232 500/pk 10x10 - UV-Visible Plastic 2.4-4.5ml non-masked 280-1100 10 15
035-078 100/pk 100x10 - Visible Plastic 35ml non-masked 340-900 100 15
Disposable test tubes
060-179 100/pk 12x75 - Visible Plastic 5ml non-masked 340-900 15
060-331 100/pk 12x75 - Visible Glass 5ml non-masked 340-900 15
060-335 50/pk 24x150 - Visible Glass 40ml non-masked 340-900 15

Home | Spectroscopy Products | Online Store | News
Application Notes  | Contact Us | About Us
Copyright © 2008 Techne® Incorporated