The Model 3345 Laboratory Ion/pH Meter
Request literature on this product via mail


Dual Display
Simultaneous display of primary and secondary measurements. A Switch Screen key is used to toggle the input channel priority.

Ease of Use
A simple push-button control panel, used in conjunction with the display, ensures that even inexperienced operators can achieve reliable results.

Dual Input (Rear Panel)
The Model 3345 Laboratory Ion Meter features two independent input channels for both sensor and temperature monitoring. A socket is provided for connection to an optional overhead or bench stirrer which can then be controlled directly from the instrument keypad.

Set Up Menus
Set-up menus are provided on the Model 3345 Laboratory Ion Meter to give access to all user defined options.

Languages Clear user instructions are available in six language options, (English, French, German, Italian,
 Portuguese or Spanish), operator selectable via the instrument set up mode.

Calibration
A choice of 1 to 3 point calibration is offered for pH measurement and 1 to 5 point calibration
 in Ion mode. The Model 3345 will automatically recognize 2,4,7,9 and 10pH buffers with a manual override facility.

Incremental Methods
The Model 3345 Laboratory Ion Meter is programmed to make the calculations necessary when carrying out Known and Sample Addition and Subtraction, and Multiple Sample Addition.

Sample Measurement
Icons and prompts guide the user through calibration and measurement routines.
GLP
Essential to the modern laboratory GLP protocols are fully supported. Product performance is continually monitored, with all logged results being time and date stamped, thus satisfying the requirements of Good Laboratory Practice.

RS232 / Analog Outputs
The Model 3345 can download directly to a computer or printer by the bi-directional RS232 link and provides an analogue output for connection to a chart recorder or datalogger.

Techne® also has a portable Ion Meter to provide the user with flexibility in a multitude of applications.  The Model 3205 Portable Ion Meter is idea for use in the field, on manufacturing sties and around process plants as well as in routine and research laboratory analysis.


Model 3345 Specifications

pH  
Range -2 to 19.999pH
Resolution 0.1 / 0.01 / 0.001pH
Accuracy ±0.003pH
mV  
Range -1999.9 to 1999.9mV
Resolution 0.1mV
Accuracy ±0.2mV
Concentration  
Range 1 x 10-9 to 9.99 x 109
Units Activity, ppm, %, mg/l, M or none
Resolution 3 significant digits
Accuracy ±0.003pX for monovalent ions
Measurement Methods Direct potentiometry / Incremental
Input Impedance >1013 Ohms
Number of Electrode Inputs 2
Temperature  
Range -10 to 105ºC / 14 to 221ºF
Resolution 0.1ºC / 1ºF
Accuracy ±0.5ºC / ±1ºF
ATC Range 0 to 100ºC / 32 to 212ºF
Manual Temperature Compensation Range 0 to 100ºC / 32 to 212ºF
Number of Temperature Inputs 2
Auto Buffer Recognition 2, 4, 7, 9 and 10pH buffers with manual override facility
Calibration 1, 2 or 3 point pH / 1, 2, 3, 4 or 5 point Concentration
Outputs Analog 1mV per digit / Bi-directional RS232 / hi-lo alarm outputs open collector 0.5A 50V max
Datalogger 100 results
Clock 24 hour, hour/min/sec or day of month/month/year - Leap year corrected
Alarm Points -2 to 19.999pH / ±1999.9mV / 1 x 10-9 to 9.99 x 109
Good Laboratory Practice Calibration reminder with an interval of 1 to 999 hours from last calibration. Timed printout with an interval from 1 second to 1 day
Display LCD Graphics
Power Power supply 9V AC
Size 275(l) x 240(w) x 150(d)mm
Weight 1.2Kg
Warranty 2 years parts and labor

Model 3345 Order Codes

548 101 Model 3345 Laboratory Ion Meter (US power supply; 117V) supplied with glass bodied combination electrode (924 005), swing arm electrode holder (903 200), ATC probe (027 232), English instruction manual, 4 & 7 pH buffer capsules, US power supply (021 030) and BNC to Leadless electrode cable (013 173).
548-301 Model 3345 Laboratory Ion Meter (European power supply) supplied with glass bodied combination electrode (924 005), swing arm electrode holder (903 200), ATC probe (027 232), English instruction manual, 4 & 7 pH buffer capsules, UK power supply (021 030) and BNC to Leadless electrode cable (013 173).
548-001 Model 3345 Laboratory Ion Meter (UK power supply; 240V)supplied with glass bodied combination electrode (924 005), swing arm electrode holder (903 200), ATC probe (027 232), English instruction manual, 4 & 7 pH buffer capsules, UK power supply (021 030) and BNC to Leadless electrode cable (013 173).
027 232 Temperature Probe
013 173 Electrode lead (BNC to Leadless)