Model 4200 Conductivity Meter |
|
The measuring range of the Model 4200 Conductivity Meter of 0 to
199.9mS is covered by five automatically switched ranges using a single
cell. Additionally, the in built software is able to calculate and report
results as Total Dissolved Solids (T.D.S.) in mg/l, Salinity in g/l or
Resistivity in MOhms. In T.D.S. mode the Model 4200 Conductivity Meter
allows a variable factor to be applied to the measured conductivity
value.
The unique measuring cell used on the Model 4200 Conductivity Meter is
designed to cope with the harsh environments encountered in many field
situations. Carefully configured solid state four plate construction
guarantees a linear response over a wide conductivity range, whilst minimizing
the effects of electrode fouling. |
The Model 4200 Conductivity Meter can be standardized in two ways - a
cell constant from 0.015 to 19.99 can be entered when a cell of known
constant is being used. Alternatively, a calibration can be executed using
a solution of known conductivity by entering its value via the set-up
menu. A single point calibration automatically updates and displays the
cell constant value. where measurements are to be made over a wide
conductivity range, and close accuracy is of importance, this instrument
offers two or three point calibration.
Solution conductivity varies significantly with temperature. For
results to be meaningful, they should be reported referenced to an agreed
value. The Model 4200 Conductivity Meter enables reference to three
internationally encountered options of 18,20 and 25°C.
The degree to which solutions change their conductivity with respect to
temperature (temperature coefficient) can be directly entered through the
instruments set-up menu. Values between 0 and 4.00%/°C are available.
Additionally, a "pure water" mode enacts the no-linear temperature
coefficient characteristics associated with low conductivity water. The
precision temperature sensing element built into the instruments cell will
correct for temperature effects between 0 and 100°C and will measure the
temperature displayed in either °C or °F.
The Model 4200 offers the facility to download data held in memory to a
printer or computer. These units incorporate a bi-directional infra-red
link which ensures that the waterproof characteristics of the products can
be maintained at all times. The infra-red link operates in conjunction
with the optional interface cradle,
allowing data interchange without the inconvenience of plugs and
sockets.

Model 4200 Specifications |
| Conductivity |
|
| Ranges |
5 switched ranges 0 to 199.9mS |
| Resolution |
0.1mS / 0.01mS / 0.001mS / 0.1µS / 0.01µS |
| Accuracy |
±0.5% ±2 digits |
| TDS |
|
| Ranges |
5 switched ranges 0 to 200g/l |
| Resolution |
0.1g/l / 0.01g/l / 1mg/l / 0.1mg/l / 0.01mg/l |
| Ranges |
±0.5% ±2 digits |
| Resistivity |
|
| Range |
0 to 20MOhms |
| Resolution |
0.01MOhms |
| Accuracy |
±0.5% ±2 digits |
| Temperature |
|
| Ranges |
-10 to 105°F / +14 to +220°F |
| Resolution |
0.1°C / 1°F |
| Accuracy |
±0.5°C / ±1°F |
| Salinity |
|
| Range |
0 to 99.9g/l |
| Resolution |
0.1g/l |
| Accuracy |
0 to 35 ±1 / 35 to 99.9 ±3 of reading |
| ATC Range |
0 to 100°C |
| Cell Constant |
Digitally settable 0.015 to 19.99 |
| Reference Temperature |
18, 20, 25°C |
| Temperature Coefficient |
0.00 to 4.00/°C linear |
| Probe |
4 plate configuration |
| Clock |
24 hour, hours/min/sec or day of month/month/year - leap year
corrected |
| Power |
3 AA cells |
| Size |
200(l) x 80(w) x 60(d)mm |
| Weight |
370g |
| Warranty |
2 years parts
and labor |

| 540 001 |
Model 4200 supplied with conductivity cell K=1 (540 520),
batteries and instruction manual |
| 033 174 |
Carrying Case |
| 025 139 |
Calibration Standard 10µS |
| 025 138 |
Calibration Standard 1413µS |
| 025 156 |
Calibration Standard 12.88mS |

Home | Electrochemistry
Products | Online
Store | News
Contact Us | About
Us | Distributors
Copyright © 2006 Techne® Incorporated |